Part Number Hot Search : 
SID1003 BC416 GP1U27R CS9014 X0605MK SED13 C5750 0N60B
Product Description
Full Text Search
 

To Download EDS3-126A Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
  this infrared gas analyzer measures the concentration of gas by utilizing the characteristic of gas (molecule consist- ing of different-kind atoms) that it absorbs the infrared ray of a specific wavelength. stable and accurate measurement is ensured due to use of a mass flow sensor with a high sensitivity, interference- compensating detector, etc. this analyzer is to be built in sampling equipment and optimally usable for measuring the concentrations of various gases. features 1. most suited for incorporation in equipment this analyzer can be readily built in customer? sampling equipment, because its analyzing block, signal process- ing circuit and power unit are all accommodated in a compact case. 2. high measurement accuracy and stability a single light source/double beam system, and a high- sensitivity mass flow sensor ensure accurate and stable measurement. 3. minimum influence by other gases an interference-compensating detector substantially suppresses interference by the disturbing gases coex- isting in sample gas. 4. wide dynamic range a range ratio of max. 1:20 can be secured in standard 2 ranges. range selection follows the external selection input signal. specifications power supply: 100v-240v ac, 50/60hz power supply voltage variation in running time; within 10%. power consumption: 130va max ambient temperature: -5 to +45 c ambient humidity: 95% rh or less enclosure temperature: -20 to +60 c enclosure humidity: 95% rh or less casing: zinc plated steel (special chromate) (jis: secc-c equivalent) steel color indoor-use dimensions (h w d): on table-top 176 433 229 mm mass (weight): approx. 9 kg warm-up time: about 4 hours (after power on) materials of gas-contacting parts: sample cell: sus304, chloroprene rubber infrared-ray transmitting window: caf 2 or sapphire internal tubing: fluoride ethylene resin data sheet zrc6 ndir type infrared gas analyzer gas inlet, outlet, purge gas inlet size: rc1/4, or npt1/4 female screw purge gad flow rate: 1l/min 0.5l/min purging is required when combustible and corrosive gases are contained in the atmosphere or the range of co 2 is less than 100ppm. in other cases, purging should be made as necessary. scope of delivery: analyzer 1, power fuse 2, instruction manual 1 mounting method: on table-top installation condition: install the analyzer at a place not ex- posed to direct sunlight or the radiation from a high temperature object. avoid vi- bration, and select a clean place free from corrosive and/or combustible gas- ses. if installing outdoors, provide a suit- able casing or cover to protect the ana- lyzer from wind, rain, etc. standard requirements for measuring gases: temperature: 0 to 50 c moisture: below a level where satu- ration occurs at 2 c (con- densation unallowable). dust: 0.3 m or less (recom- mended membrane filter; zbbm6) pressure: 10kpa or less (flow rate: 0.5l/min)(gas outlet side should be open to the at- mospheric air) standard control for sample gas: calibration gas: dry gas interfere control gas: 2 c saturation measuring system: infrared-ray absorption method, non-dis- persion, differential flow system, single light source, double-beam system. date EDS3-126A apr. 1, 2011
zrc6 2 code symbols description measuring components co (carbon monoxide) co 2 (carbon dioxide) no (nitric oxide) so 2 (sulfur dioxide) basic structure single-component analyzer (bench) b d p a 6 1st measuring range note: for measuring range, refer to table. 0 to 50ppm 0 to 100ppm 0 to 200ppm 0 to 250ppm 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% 2nd measuring range 1st range 0 2 2.5 4 5 8 10 20 1y00f rc6 z-- 12345678 10111213 1415 9 gas aperture shape rc1/4 npt 1/4 female screw a b power supply 100v ac to 240v ac, 50/60hz measurable gas type atmosphere combustion exhaustion gas converter exhaustion gas 0 e f g a b c d e f g h j k l m n p r 0 2 3 4 5 8 1 9 principle diagram of infrared type measurement sample gas outlet ac amplification sample cell sample gas inlet t rimmer reference cell interference compensation detector distribution cell (interference filter) infrared light source (single-beam light source) motor rotation sector sector sync signal output 4 to 20ma dc 0 to 1v dc (simultaneous output) range switching synchronous genarator smooth- ing dc amplification converter linearizer v/i measurable components and measuring ranges: measurable components co co 2 no so 2 carbon monoxide carbon dioxide nitric oxide sulfur dioxide 0 to 100ppm minimum measuring range 0 to 50ppm 0 to 100ppm 0 to 100ppm 0 to 100vol % maximum measuring range 0 to 100vol % 0 to 2000ppm 0 to 10vol % measuring ranges: shown on another table range selection: short circuit between external terminal 13 and14 switches to high range side. output signal: 0 to 1v dc/4 to 20ma dc simultaneous output linearity: 2% of full scale 0 to 1v dc/permissible load resistance; 100k ? or more 4 to 20ma dc/permissible load resis- tance; 550 ? or less repeatability: 0.5% of full scale zero drift: 2% of full scale/week span drift: 2% of full scale/week response time: maximum within 20 seconds including gas replacement time of sample gas. (for 90% fs response) but differs from the length of sample cell. measured gas flow rate: standard 0.5l/min 0.1l/min
3 table 1. measurable component and range availability check table (1) single-component analyzer (no, so 2 , co 2 , co) 2nd range 1st range e 0 to 500ppm c0 to 200ppm e f 0 to 1000ppm g 0 to 2000ppm h 0 to 5000ppm j 0 to 1% k 0 to 2% 0 to 500ppm f 0 to 1000ppm g 0 to 2000ppm d0 to 250ppm h 0 to 5000ppm j0 to 1% k0 to 2% l0 to 5% m0 to 10% n0 to 20% p0 to 50% r0 to 100% a0 to 50ppm b 0 to 100ppm b 0 to 100ppm c 0 to 200ppm d 0 to 250ppm : no analyzer measurable range : so 2 analyzer measurable range : co 2 analyzer measurable range : co analyzer measurable range 2nd range 1st range c0 to 200ppm e l 0 to 5% m 0 to 10% n 0 to 20% p 0 to 50% 0 to 500ppm f 0 to 1000ppm g 0 to 2000ppm d0 to 250ppm h 0 to 5000ppm j0 to 1% k0 to 2% l0 to 5% m0 to 10% n0 to 20% p0 to 50% r0 to 100% a0 to 50ppm b 0 to 100ppm r 0 to 100%
1 2 3 4 5 6 78 9 10 11 12 13 14 15 16 source 2a outlet purge inlet span cal. (vr4) zero cal. (vr7) fuse (229) (176) (164) connecting diagram output 1 (0 to 1v dc) output 2 (4 to 20ma dc) power supply 100v ac to 240v ac range check (0.04 to 0.2v dc) purge gas inlet rc1/4 or npt1/4 connecting terminal sampling gas inlet rc1/4 or npt1/4 sampling gas outlet rc1/4 or npt1/4 fuse zero adj./span adj. < front > < upper > < side > < back > (210) (432.5) outline diagram (unit:mm) < analyzer main unit > zrc6 printed in japan caution on safety *before using this product, be sure to read its instruction manual in advance. international sales div sales group gate city ohsaki, east tower, 11-2, osaki 1-chome, shinagawa-ku, tokyo 141-0032, japan http://www.fujielectric.com phone: 81-3-5435-7280, 7281 fax: 81-3-5435-7425 http://www .fjielectric.com/products/instruments/ information in this catalog is subject to change without notice.


▲Up To Search▲   

 
Price & Availability of EDS3-126A

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X